The Evaluation of Noviplex Plasma Prep Cards for Regulated Bioanalysis
L. F. Hvozda1, C. D. Christianson1, T. Schlabach2
1Alturas Analytics, Inc., 2Novilytic LLC

Purpose
The purpose of this work is to determine if the Noviplex plasma prep cards produce the precision required to perform regulated bioanalysis.

Methods
Human whole blood was spiked with Indomethican at three concentrations; 9.00, 70.0 and 1,200 ng/mL. Two different types of Noviplex cards at n=6 were then spotted with the whole blood: the original Noviplex (40 µL blood volume per card) and the Noviplex DUO (60 µL blood volume per card). Once spotted the cards were allowed to dry for at least 15 minutes. The cards are designed to remove red blood cells from whole blood using capillary action to pull the matrix through the card. The resulting plasma spreads out onto the plasma discs and dries. The plasma is split onto two separate plasma disc when the blood is processed on the DUO cards.The plasma discs were then placed into a 96 DWP and extracted with methanol. The extracts were analyzed on an ABSciex 5500® QTRAP coupled with a Shimadzu HPLC system with a flowrate of 0.7 µUmin. The column used for the binary gradient analysis was a Supelco HSC 18 heated to 50°C (3 µm, 50 mm X 2.1 mm). The data collected from the extracts was quantified against a calibration curve extracted from plasma with a dynamic range of 1-1500 ng/mL.

Results
The data indicates that that the plasma aliquot dried onto the disc, extracted and analyzed resulted in a CV that exceeds GLP specifications. The CV at 9.00, 70.0 and 1200 ng/mL was found to be 11.6, 7.4 and 5.8 respectively. The accuracy at 70.0 and 1200 ng/mL was found to be 3.1 and 2.4 %bias from nominal concentrations respectively. The low concentration of 9.00 ng/mL accuracy was found to be 27.8 %bias from nominal which indicates that direct quantitation from a standard curve prepared from plasma will not provide the required accuracy at the lower limits of quantitation. The standard curve must also be prepared on the cards in order to compensate for the difference in extraction efficiency or matrix effects that may be occurring at the lower concentrations.

The data generated from the Noviplex DUO card by extracting the left and right plasma spot separately indicates that the plasma distribution between the two spots are equivalent. The percent difference between the left and right spot for the 9.00, 70.0 and 1200 ng/mL was found to be 0.9%, 5.3% and 0.6% respectively.

Conclusion
The Noviplex plasma prep cards provide precise plasma spot data that when extracted and analyzed may be used for regulated bioanalysis.

Please complete the form below in order to access our resource:

"*" indicates required fields